Chlorine 3017M DPD chlorine analyzer

Chlorine 3017M DPD chlorine analyzer

More Pictures

  • Chlorine 3017M DPD chlorine analyzer
  • WTW-860151-Chlorine_3017M-DPD_chlorine_analyzer-front-open

Chlorine 3017M DPD chlorine analyzer

5312 In stock
SKU: 860151
€5,312.00
€5,312.00
OR
OR
Description: online analyzer for photometric measurement of free and total chlorine, according to colorimetric DPD Method (ISO &US EPA); outputs (selectable): 4 ÷ 20 mA or RS 485 Modbus; range: 0 ÷ 5 mg/l; sample inlet device not included;
Measurement principle:  colorimetric with N, N-Diethyl-p-phenylenediamine (DPD)
Parameter:  free and total chlorine
Range:  0 ÷ 5 mg/l, Cl
Resolution:  0.0001 NTU
Accuracy:  ±0.03 mg/l or ±5%, whichever is greater
Limit of detection:  0.03 mg/l
Measurement interval:  programmable; 2.5 ... 60 minutes
Sample temperature:  5 ÷ 45°C (41 ÷ 113°C)
Sample flow rate:  50 ÷ 1000 ml/min. when using sample inlet device
Inlet pressure:  0.07 ÷ 1.40 bar (1 ÷ 20 psi) with Sample Inlet Device
Reagent consumption:  ~30 days per bottle at a 2.5 minute measurement interval
Calibration:  factory calibrated, 1-point if required
Display:  2.8 x 6 cm backlit LCD
Mounting:  4 mounting struts bolted to back of unit
Regulatory compliance:  US EPA regulations 40 CFR 141.74 and 40 CFR 136.3;
Standard method 4500-CL-G; 
US EPA method 334.0; 
ISO method 7393-2
Reagent storage life:  before hydration: buffer and indicator: 5 years. DPD powder: 1 year
after hydration: ~30 ÷ 40 days
Power:  115 ÷ 230VAC, 50–60 Hz
Relays: 2 x 6A/30 VDC
Current outputs:  1 x 4 ÷ 20 mA 
Digital output:
RS-485, Modbus RTU
Light source: 
Class 1 LED; wavelength centered at 525 nm (light path length: >1cm)
Ambient temperature:
5 ÷ 55°C (41 ÷ 131°F)
Protection class: 
IP66
Material:  enclosure: polycarbonate
flow cell: CPVC
terminal block: polyamide
fan assembly: ABS
reagent pump assembly: cyanoacrylate body and stainless steel rollers
sample pump assembly: PA12 is polyamide12 and POM is polyacetal
Weight:  8kg (18Ibs)
Warranty:  2 years 

Accessories