LR 325/001 sensor

LR 325/001 sensor

LR 325/001 sensor

0 In stock
€0.00
application:

conductivity measurement ulra pure water,  designed to work with EcoLine LF170, QuadroLine LF296, S182, S184, 2020XT controllers

measuring principle: conductivity sensor with 2-electrode measuring cell, K = 0.01 cm-1
measuring range: 0.0001 µS/cm ÷ 30 µS/cm
process temperature: 0 ÷ 100ºC (integrated NTC sensor, ± 0.2 K) 
pressure:  2 bar 
material: stainless st. 1.4571 / POM
dimensions: ∅12 x 165mm
cable: 1.5m
ordering code: 301962 + adapter ADA/AMPH-LAB-LF (303212) for connection to Ecoline LF170 or KI/LF-MIQ/S (505543) for connection to S182, S184, 2020XT controllers
protection class: IP68